CardioPerfect Workstation. Resting ECG Module - User Manual. DIR 80012333 Ver. Before installing the Resting ECG software. In this chapter, you will find information about the structure of the Welch Allyn CardioPerfect resting ECG software. Welch Allyn Cardioperfect Workstation, Software Version 1.6.5 - Installation Guide User Manual. Restore the database. Welch Allyn Equipment Manuals Directory ManualsDir.com - online owner manuals library.
Page 46: Hardware, Install unilink or cpcom, Cardioperfect workstation, Resting ecg module - user manual
Advertising
CardioPerfect Workstation
Resting ECG Module - User Manual
DIR 80012333 Ver. E
46 / 54
11. Before installing the Resting ECG software
In this chapter, you will find information about the structure of the Welch Allyn CardioPerfect resting ECG software.
11.1. Structure
The workstation consists of two elements:
Hardware: the recorder and the interface to the computer.
Welch Allyn Cardioperfect Manual
Software: the Resting ECG module that runs in the Welch Allyn CardioPerfect Workstation.
The recorder must be connected to the computer that runs the module.
Software
Welch Allyn Cardioperfect Workstation Software
The resting ECG module is automatically installed with Welch Allyn CardioPerfect Workstation. For instructions on how to install and configure Welch Allyn CardioPerfect Workstation, please refer to the Installation manual.
Hardware
Three different types of recorders are available, the Portable recorder, the MD recorder and the Pro recorder. The recorder with a patient cable, with which you record the ECG, is connected to the PC through an interface; different types of interfaces are available:
The Portable- or MD recorder communicates with the computer through an optical fiber which is connected to a so-called Unilink (USB or RS232) or to a so-called CPCOM card.
The Pro recorder uses a Prolink USB or Prolink RS232 to communicate with the computer.
Install Unilink or CPCOM?
A Unilink is a small optical to electrical interface. There are two versions of the Unilink: the serial or RS232 Unilink and the USB Unilink.
The Welch Allyn CardioPerfect ECG recording system can optionally be supplied with a CPCOM card. This is an expansion card which plugs into a PCI expansion slot. This card essentially takes over the function of the Unilink but, compared to the RS232 version, it has an additional capability to buffer the incoming data.
Advertising
Popular Brands
Popular manuals
Advertising
CardioPerfect Workstation
Resting ECG Module - User Manual
DIR 80012333 Ver. E
17 / 54
Leadoff, Framing Errors and DCOffset conditions:
If the application detects a Lead-off, DC offset, or Framing error within 10 seconds after you have started an ECG recording, the recording automatically restarts.
Welch Allyn Cardioperfect Workstation User Manual Guide
In the case of a Leadoff condition,
the ECG recording will not start until the signal is stabilized. The progress bar of the recording will remain at 0 until all the signals are of acceptable quality. You can use the override button to disable the automatic restart. DCOffSet
The DCOFFSET DETECTED! message displays in the progress bar when
Welch Allyn Software
battery power is interrupted or when there is a spike in voltage.
Framing Errors Framing errors display in the progress bar and can occur in CPWS
configurations when the Pro recorder is connected via a serial cable and the
packets of information are delayed.
Leadoff
The Leadoff message displays in the workspace when an electrode is
disconnected. The real-time monitor will displ
ay “Bad signal on…”. The trace
Welch Allyn Cardioperfect Workstation User Manual 10
shows square waves until the signal is stabilized.